Source container FQG60

Hauser

15

Installation

Incoming acceptance, 

transport

The radiation source container also serves as Type-A packaging (IATA rules) for the radiation source. 

For transport, it is protected in the box by foam packaging.

Package dimensions: 375 x 330 x 275 mm (14.8 x 13 x 10.8 in)

CAUTION

!

Don’t use bow-type handles for transportation

Only use the bow-type handles (3) of the shutter to remove the source container from the 

packaging by hand. 

To transport the source container, use the slots on the fastening lug, e.g. with a lifting eye. 

A0018493

1

Lifting  eye

2

Fastening lug

3

Handles

Mounting hints

The radiation source container can be mounted in one of the following ways:

• Using L-profiles or a mounting plate (not pressurized and not in contact with the process) directly 

on the vessel or pipe (

ä

18).

• On an external construction with low to zero vibration.

• Directly on the pipe at the customer's site using the clamping device FHG61 (

ä

29).

CAUTION

!

Mounting of source container

All maintenance such as mounting, removal or replacement of the radioactive source may only be 

carried out by supervised personnel who have been specially trained in radiation procedures 

according to local regulations or the handling permit. Ensure that this is allowed in the handling 

permit. Local conditions are to be observed.

All work must be carried out as quickly as possible and from a distance as large as possible 

(shielding!). Safety procedures (e.g. blocking of access) must also be carried out to protect 

personnel from all possible risks.

Mounting and dismounting is only allowed in the "OFF" position, secured by the padlock..

Take into account the weight of the radiation source container: max. 18 kg (39.69 lbs).

To ensure the correct operation of the switch-ON/switch-OFF function, no parts of the vessel, pipe 

and clamping device may project into the area of the shutter. If the unit is secured via the mounting 

holes ø11 mm (0.43 in) this should not warp or damage the metal housing.

If using the device in non-stationary systems, appropriate measures must be taken to ensure the 

device cannot be lost, and to protect it from collision and impact.

If using methods other than the mounting plate or L-profiles to secure the device, we recommend 

the use of the clamping device FHG61.

Mounting instructions are provided in the documentation: 

SD00330F/00 and SD0331F/00.

1

2

O

FF

O

FF

O

N

O

N

RADI

O

ACTIV

E

RADI

O

ACTIV

E

3

The foam packaging can be disposed like ordinary consumer waste.

Summary of Contents for FQG60

Page 1: ...FQG66 devices areble for larger activities In the case of density measurement it is suitable for pipe outer diameters ranging from 48 to 273 mm 1 89 to 10 7 in Your benefits Small size lightweight de...

Page 2: ...Source container FQG60 2 Endress Hauser A0023555 Order code Ext ord cd Ser no www endress com deviceviewer Endress Hauser Operations App XXXXXXXXXXXX XXXXX XXXXXX XXX XXXX XX Serial number 1 3 2...

Page 3: ...6 Mounting position for level limit detection 17 Mounting device supplied by customer 18 Torque for the mounting screws supplied by customer 20 Post installation check 20 Operation 21 Safety instructi...

Page 4: ...he device against extreme influences e g chemical products weather mechanical impacts vibrations when operated or stored Always secure the OFF position using the padlock Before switching ON the radiat...

Page 5: ...mma radiation Application for the permit must be made to the Land government or the authority responsible Land Offices for Environmental Protection Trade Inspection Offices etc The Endress Hauser Sale...

Page 6: ...can result in serious or fatal injury A0011191 DE CAUTION This symbol alerts you to a dangerous situation Failure to avoid this situation can result in minor or medium injury A0011192 DE NOTICE This s...

Page 7: ...h do not take into account production dependent fluctuations of the source activity and tolerances of the measuring devices Maximum activity of the radiation source 137 Cs 1 11 GBq 30 mCi CAUTION The...

Page 8: ...240 Emission Angle Application Option model 5 40 deg level 90 30 60 120 300 330 0 270 210 240 180 150 40 60 80 100 Option model in Feature 100 Prepared for Source Activity Activity in MBq max value 1...

Page 9: ...ivity in MBq max value 100 in Sv h AC 18 5 0 15 AD 37 0 29 AE 74 0 59 AF 111 0 88 AG 185 1 47 AH 370 2 94 AK 740 5 87 AL 1110 8 81 RS 0 74 0 01 90 30 60 120 300 330 0 270 210 240 180 150 40 60 80 100...

Page 10: ...flat gasket plate into the radiation path 21 Nameplate of source container 14 Radiation emission channel A0018392 A Feature 240 Emission Angle Application Option model 3 20 deg limit switch density B...

Page 11: ...0 deg limit switch density 20 radiation emission angle Level measurement A0018489 Dimensions mm in Feature 240 Emission Angle Application Option model 5 40 deg level 40 radiation emission angle OFF OF...

Page 12: ...ce insert for theft protection OFF OFF ON ON 61 2 4 244 9 61 132 5 2 180 7 09 170 6 69 232 9 13 349 13 7 52 2 05 11 0 43 Component Material Source insert and internal components Stainless steel 304 1...

Page 13: ...ure 40 to 120 C 40 to 248 F Ambient pressure Atmospheric pressure Vibration resistance IEC EN 60068 2 64 test Fh 10 to 2000 Hz 0 01 g2 Hz Shock IEC 60068 2 27 test Ea 30 g 18 ms 3 shocks direction axi...

Page 14: ...ion source 9 Marking Hochradioaktive Strahlenquelle according to German regulations if required 10 137 Cs 11 Serial number of the source capsule provided for source tracking if required 12 Activity in...

Page 15: ...e been specially trained in radiation procedures according to local regulations or the handling permit Ensure that this is allowed in the handling permit Local conditions are to be observed All work m...

Page 16: ...st be mounted at the height of or slightly above the maximum level for continuous level measurement The radiation must be aligned exactly with the compact transmitter mounted opposite The source conta...

Page 17: ...arge measuring ranges The use of several sources can be necessary not only from the aspect of large measuring ranges but also for accuracy reasons 2 1 1 EIN ON EIN ON EIN ON EIN ON 4 0 4 0 A0018505 1...

Page 18: ...ounting holes 11 mm 0 43 in may be used for this purpose CAUTION The intermediate space must be blocked off if necessary to limit access A0018506 Dimensions mm in A Mounting plate A0018507 Dimensions...

Page 19: ...ty measurement on horizontal pipes With this type of orientation it is advisable to mount the FQG60 above the pipe This prevents the buildup of solid particles or liquids on the shutter However attent...

Page 20: ...Behavior in event of empty process vessel or pipe CAUTION Radiation Once the unit has been correctly mounted the control area of the empty process vessel has to be measured If it is necessary this are...

Page 21: ...Option model C Remove padlock in OFF position Feature 020 Option model B with snap hook Remove padlock in OFF position A0018513 Using the bow type handles move the shutter lead shielding from the OFF...

Page 22: ...ut 21 After commissioning a density measurement the calibration plate is introduced into the radiation path under constant conditions as described below and the displayed density value at the FMG60 is...

Page 23: ...t point 10 is available with the Gammapilot M device for recalibration This point can be entered if the measuring conditions have changed for example due to buildup in the measuring tube Io correspond...

Page 24: ...utine repair or maintenance must be performed by the gauge manufacturer or distributor or in the USA by a person specially authorized by NRC or an Agreement State Measures in case of corrosion If cons...

Page 25: ...all be performed by a person or an organization authorized to provide leak test services or using a leak test kit Leak test kits shall be used according to its supplier s instructions Records of the l...

Page 26: ...sition In this case the radiation source container must be unbolted from its mounting position CAUTION Dismounting Point the emission channel towards a very thick wall e g steel or lead or mount a thi...

Page 27: ...rried out to protect personnel from all possible risks The source container may only be disassembled in OFF position Make sure that the OFF position is secured with a padlock Return Federal Republic o...

Page 28: ...requires the source container be delivered first and if the source must be delivered subsequently transportation drums are used for shipping Other countries We can only ship radioactive sources once...

Page 29: ...rt right hand column Configure the selected product The Product Configurator for the selected product opens From your Endress Hauser Sales Center www addresses endress com B A Product Configurator the...

Page 30: ...er diameter 48 to 77 mm 1 89 to 3 03 in Gammapilot M FMG60 TI00363F 00 Technical Information for Gammapilot M FMG60 BA00236F 00 Operating Instructions for Gammapilot FMG60 HART BA00329F 00 Operating I...

Page 31: ...Source container FQG60 Endress Hauser 31 Manufacturer Declaration Radiation Source Container A0037353...

Page 32: ...www addresses endress com 71481438...

Reviews: