Operation
Deltabar S PROFIBUS PA
28
Hauser
6.5 Locking/unlocking operation
Once you have entered all the parameters, you can lock your entries against unauthorised and
undesired access.
You have the following possibilities for locking/unlocking the operation:
• Via a DIP-switch on the electronic insert (
• Via the on-site display (optional)
• Via digital communication.
The
-symbol on the on-site display indicates that operation is locked. Parameters which refer
to how the display appears, e.g. LANGUAGE and DISPLAY CONTRAST can still be altered.
• If operation is locked by means of the DIP-switch, you can only unlock operation again by
means of the DIP-switch. If operation is locked by means of remote operation e.g. FieldCare,
you can only unlock operation again by means of remote operation.
The table provides an overview of the locking functions:
Locking via
View/
read
parameter
Modify/write via
1)
1)
Parameters which refer to how the display appears, e.g. LANGUAGE and DISPLAY CONTRAST can still be altered.
Unlocking via
On-site
display
Remote
operation
DIP-Switch
On-site
display
Remote
operation
DIP-Switch
yes
no
no
yes
no
no
On-site display
yes
no
no
no
yes
yes
Remote operation yes
no
no
no
yes
yes
Locking/Unlocking operation via on-site display or remote operation
Locking operation
1. Select INSERT PIN NO. parameter,
Menu path: GROUP SELECTION
OPERATING MENU
OPERATION
INSERT PIN
NO.
2. To lock operation, enter "0" for this parameter.
Unlocking operation
1. Select INSERT PIN NO. parameter.
2. To unlock operation, enter "2457" for the parameter.