Deltabar FMD72
Description of device parameters
95
HP
LP
LP
HP
L
T
P
PV
L
T
HP
LP
+
-
t
P
Processing
sensor HP
measurement
Processing
sensor LP
measurement
Simulation pressure
Simulation pressure
Processing
differential press.
Corrected
press.
Measured
pressure
pressure
Simulation
differential pressure
Calib. offset
+ pressure
inversion
Pressure
simulation
Measured
differential
pressure
Processing
level linear
Level before
linearization
Tank
content
Processing
Tank content
(lin.tab)
Measuring mode level only
Simulation level
Simulation tank content
Measured
value
Measured
differential pressure
Level before linearization
Tank content
Measuring mode
Lin. mode
Processing current output
Current
output
Diagnosis/
Device status
Sim. current
Sim. Error no.
Measured
A0017376-EN
Sim. diff.press. (113)
Navigation
Diagnosis
®
Simulation
®
Sim.diff.press. (113)
Write permission
Operators/Service engineers/Expert
Description
Enter the simulation value. See also "Simulation Mode".
Prerequisite
"Simulation mode" = Differential pressure
Value at switch-on
Current differential pressure measured value
Summary of Contents for Deltabar FMD72
Page 2: ......
Page 107: ......
Page 108: ...www endress com worldwide 71191239 71191239 BA01044P 00 EN 02 12 71191239 CCS COSIMA...