Deltabar FMD71, FMD72
Description of device parameters
47
Navigation
Expert → Measurement → Sensor Trim HP → Hi Trim Measured (130)
Write permission
No write permissions. Parameter is read only.
Description
Displays the serial number of the sensor module HP (11 alphanumeric characters).
Lo trim sensor (131)
Navigation
Expert → Measurement → Sensor Trim HP → Lo Trim Sensor (131)
Write permission
No write permissions. Parameter is read only.
Description
Sensor module recalibration by entering a target pressure while simultaneously and
automatically accepting a reference pressure present for the lower calibration point.
Hi trim sensor (132)
Navigation
Expert → Measurement → Sensor Trim HP → Hi Trim Sensor (132)
Write permission
No write permissions. Parameter is read only.
Description
Sensor module recalibration by entering a target pressure while simultaneously and
automatically accepting a reference pressure present for the upper calibration point.
Start current (134)
Navigation
Expert → Output → Current output → Startcurrent (134)
Write permission
Service engineers/Expert
Description
Entry of the start current. This setting also applies in HART multidrop mode.
Options
• 12 mA
• Max Alarm (22 mA, cannot be adjusted)
Factory setting
12 mA
Curr. trim 4mA (135)
Navigation
Expert → Output → Current output → Curr. trim 4mA (135)
Write permission
Service engineers/Expert