CNGmass DCI
Commissioning
Hauser
47
!
Note!
Quick Setup automatically deactivates the function EMPTY PIPE DETECTION (6420) so that the
device can measure flow at low gas pressures.
Quick Setup "Gas measurement"
HOME position
→
F
→
MEASURED VARIABLE (A)
MEASURED VARIABLE
→
O
→
QUICK SETUP (B)
QUICK SETUP
→
N
→
QS-GAS MEASUREMENT (1004)
Function No.
Function name
Setting to be selected (
P
)
(to next function with
F
)
1004
QS GAS MEASUREMENT
YES
After
F
is pressed by way of confirmation, the Quick
Setup menu calls up all the subsequent functions in
succession.
Æ
6400
ASSIGN LOW FLOW CUTOFF
On account of the low mass flow involved when gas
flows are measured, it is advisable not use a low flow
cutoff.
Setting: OFF
6402
ON-VALUE LOW FLOW CUT OFF
If the ASSIGNMENT LOW FLOW CUTOFF function
was not set to "OFF", the following applies:
Setting:
0.0000 [unit]
User input:
Flow rates for gas measurements are low, so the
value for the switch-on point (= low flow cutoff)
must be correspondingly low.
6403
OFF-VALUE LOW FLOW CUTOFF
If the ASSIGNMENT LOW FLOW CUTOFF function
was not set to "OFF", the following applies:
Setting:
50%
User input:
Enter the switch-off point as a positive hysteresis in
%, referenced to the switch-on point.
Æ
Back to the HOME position:
→
Press and hold down Esc key
X
for longer than three seconds or
→
Repeatedly press and release Esc key
X
→
Exit the function matrix step by step