Installation
Cleanfit W CPA450
10
Hauser
3.2.2
Installation notes
Suitable sensors
The following sensors are suitable for installation in the CPA450:
• Digital sensors with Memosens technology, length 120 mm / 4.72"
• pH/ORP glass electrodes, length 120 mm / 4.72"
• ISFET sensors: only sensors listed in "Accessories" chapter
• DO sensors, length 120 mm / 4.72"
• For the conductivity sensor CLS15 a modification is available
Installation positions
The permissible installation angle of the assembly depends on the sensor:
• Glass electrodes and digital sensors with Memosens technology:
Install the assembly at an angle of at least 15° from the horizontal (see Fig. 3).
• ISFET sensors:
When using an ISFET sensor, there are, in principle, no restrictions to the installation. An
installation angle between 0° and 180° is, however, recommended.
a0005327
Fig. 3:
Installation angle of the assembly
A
Glass sensors: 15° to horizontal
B
ISFET sensors 0 to 180° recommended
Install the assembly so that the sensor is kept wet at all times.
15°
15°
A
B