Cleanfit P CPA472D
Installation
Hauser
9
a0006094
Fig. 3:
Installation examples with recommended installation angle (glass electrodes)
A
Tank
B
Flow assembly DN 50/80
C
Horizontal pipe minimum DN 80
D
Ascending pipe minimum DN 80
"
Caution!
• Avoid a siphon effect
1)
at the rinse chamber outlet when installing with inclined orientation. The
inlet to the rinse chamber must be from below.
!
Note!
• The minimum diameter for direct installation in pipework is DN 80. This diameter is required so
that the assembly has sufficient distance from the pipe wall when brought into the "Measuring"
position.
• Use a flow assembly to install the assembly in smaller pipe diameters (see Accessories).
• When designing the installation nozzle, please take into account the total immersion depth in
operation (sensor holder not inserted). Ensure that the sensor is always immersed in the medium
during operation (see "Dimensions").
> 15 °
> 15 °
> 15 °
A
B
C
D
1)
Siphon effect: line emptied by vacuum