Installation
Cleanfit P CPA472D
16
Hauser
3.4
Sensor installation
3.4.1
Preparing sensor and assembly
"
Caution!
If turned in the opposite direction, the stop lock bolt does not engage. This could, however, loosen
the sensor holder. The reason for this is adhesions on the lower part of the sensor holder. These can
cause the sensor holder to get stuck, producing a counterforce when unscrewing the sensor holder.
1. Remove the protection cap from the sensor.
Make sure the sensor shaft is fitted with the
O-ring and the thrust collar (Fig. 13).
2. Moisten the sensor shaft before installing the
sensor.
3. Depending on the assembly version:
a.
Manually driven assembly:
Pull the retractable pipe as far as possible
out of the assembly.
b.
Pneumatically driven assembly:
Move the assembly into the "Service"
position.
4. Turn the stop lock bolt through 90° so that
the plastic grooves are located above the
recesses (Fig. 14, A).
5. Turn the retractable pipe
clockwise
until
the stop lock bolt engages (B).
a0006214
Fig. 13:
Sensor installation
1
Thrust collar and O-ring
a0002738
Fig. 14:
Stop bolt
1
A
B