Cleanfit P CPA472
Technical data
Hauser
37
8
Technical data
8.1
Environment
8.2
Process
"
Caution!
The process pressure for a manually driven assembly must not exceed 4 bar (58 psi)!
Ambient temperature
Ambient temperature not below 0 °C (32 °F).
The maximum permissible temperature for electric limit position switches (NAMUR type)
is 90 °C (194 °F).
Process pressure
0 to max. 4 bar (0 to max. 58 psi) overpressure for manual operation
0 to 6 bar (0 to 87 psi) overpressure at 20 °C (68 °F) for pneumatic operation
Process temperature
0 to 80 °C (32 to 176 °F) in PVDF
0 to 60 °C (32 to 140 °F) in PP
Pressure temperature diagram
a0008346
Fig. 26:
Pressure temperature diagram
A
PVDF
B
PP
Flow velocity
a0008347
Fig. 27:
Permissible medium velocity in m/s (ft/s) depending on the medium temperature in °C (°F)
p [bar]
T[°C]
A
B
20
80
60
4
40
T[°F]
68
176
140
104
p [psi]
58
6
87
0
32
0
0
PVDF
PP
6.56
3.28
[ft/s]
0.33
68
176
140
T [°F]
[m/s]
2
1
0.1
20
80
60
T [°C]