Commissioning
Cerabar PMC71B HART
24
Hauser
4. Enter the appropriate value in each parameter or select the appropriate option. These
values are written directly to the device.
5. Click "Next" to go to the next page.
6. Once all the pages have been completed, click "End" to close the
Commissioning
wizard.
If the
Commissioning
wizard is cancelled before all the necessary parameters have been
configured, the device may be in an undefined state. In such situations, it is advisable to
reset the device to the factory default settings.
Example: Outputting of the pressure value at the current output
Pressure and temperature units are converted automatically. Other units are not
converted.
In the following example, the pressure value should be measured in a tank and output on the
current output. The maximum pressure of 450 mbar (6.75 psi) corresponds to the 20 mA
current. The 4 mA current corresponds to a pressure of 50 mbar (0.75 psi).
Prerequisites:
• Measured variable in direct proportion to the pressure
• Due to the orientation of the device, there may be pressure shifts in the measured value
(when the vessel is empty or partly filled, the measured value is not zero)
Perform a position adjustment if necessary
• In the
Assign PV
parameter, the
Pressure
option must be selected (factory setting)
Display: In the
Guidance
menu
Commissioning
wizard, keep pressing the key until you
reach the
Assign PV
parameter. Press the key to confirm, select the
Pressure
option and
press to confirm.
B
450 mbar
20 mA
A
50 mbar
4 mA
I
[mA]
20
4
50
p
[mbar]
450
A0039009
A
Lower range value output
B
Upper range value output