Description of device parameters
This section describes the parameters in the order they are arranged in the "Expert"
operating menu.
13.1 Expert → System
Operator code (021)
Write permission
Operators/Service engineers/Expert
Description
Use this function to enter a code to lock or unlock operation.
User entry
• To lock: Enter a number ≠ the release code (value range: 1 to 9999).
• To unlock: Enter the release code.
Note
The release code is "0" in the order configuration. Another release code can be defined in
the "Code definition" (023) parameter. If the user has forgotten the release code, the
release code can be visible by entering the number "5864".
Factory setting
0
13.2 Expert → System → Instrument info
Device tag (022)
Navigation
Setup → Extended setup → Device tag
Write permission
Operator/Maintenance/Expert
Description
Enter the device tag, e.g. TAG number (max. 32 alphanumeric characters).
Serial number (096)
Write permission
Parameter is read only. Only Hauser Service has write permission.
Description
Displays the serial number of the device (11 alphanumeric characters).
Firmware version (095)
Write permission
No write permissions. Parameter is read only.