Operation
multi EA 4000
70
6
Operation
6.1
Activation / warm-up phase / setup
CAUTION
Crushing hazard at the autosampler FPG 48! The autosampler arm, the gripper and
the boat carousel move during activation and initialization. Keep an adequate distance
to avoid having your hands crushed.
NOTICE
In case of large temperature differences between storage and the installation wait un-
til the multi EA 4000 has adapted to the ambient temperature before connecting it in
order to prevent damage to the device from condensation water.
Before switching on check the following:
The gases oxygen and argon (for Cl determination or the determination of elemen-
tary carbon) are connected with an inlet pressure of 200 – 400 kPa (29 – 58 psi).
A combustion tube has been inserted in the furnace.
The gas hoses are connected to the gas sluice or the TIC reactor.
The analysis modules have been connected.
The manual sample feed or the autosampler have been positioned and adjusted.
The PC is connected and the multiWin software installed.
1.
Open the gas flows at the pressure reducer (inlet pressure 200 - 400 kPa or 29 -
58 psi).
2.
Switch on the PC.
3.
Switch on the sampler FPG 48 (optional).
The sampler FPG 48 initializes.
4.
Switch on the basic device multi EA 4000 from the mains switch at the rear of the
device and the On/Off switch at the front (behind the doors).
5.
Switch on the analysis modules.
The devices are booted. The status lights at the front panels of the basic de-
vice and the modules illuminate.
6.
Start the multiWin program.
7.
Log in as user.
Enter A
DMIN
as U
SER
and as P
ASSWORD
. This user has been installed during initial
program installation and has administrator rights.
8.
Click the [I
NITIALIZE ANALYZER
] button.
The initialization and the automatic detection of all connected and switched
on components is performed after successful login.
Switching on sequence
Summary of Contents for Analytik Jena multi EA 4000
Page 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Page 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Page 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Page 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Page 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Page 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Page 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Page 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...