Function and setup
multi EA 4000
36
3.6
Manual TIC solids module
The manual TIC solids module allows the determination of TIC in solids in conjunction
with the multi EA 4000 C or the multi EA 4000 C/S. Acidification of the samples and
purging of TIC is performed at elevated temperatures. The measuring gas is trans-
ported with the carrier gas to the detector of the C/S module. The manual TIC solids
module is designed exclusively for the manual supply of samples only. Automated
sample supply with an autosampler is not possible.
1 Manual TIC solids module
2 C/S module
3 Basic device multi EA 4000
Figure 26 Manual TIC solids module
3.6.1
Technical data
General characteristics
Designation/type
Manual TIC solids module
Dimensions
300 x 750 x 550 mm
Mass
approx. 10 kg
Procedural data
Measuring methods
TIC
Detection
NDIR (coupled with VITA method)
Sample volume
up to 3000 mg
Gas supply
Oxygen min. 2.5
200 – 400 kPa (29 – 58 psi)
Electrical variables
connection
220-240 V AC (±10 %), 50/60 Hz
(100 V AC / 115 V AC upon request)
Fuses
T 10 A H (Only use original fuses by Analytik
Jena AG!)
Maximum power consumption
415 VA
Summary of Contents for Analytik Jena multi EA 4000
Page 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Page 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Page 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Page 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Page 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Page 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Page 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Page 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...