ENDEVCO MODEL 136 DC AMPLIFIER
INSTRUCTION MANUAL
IM136
Page 18
After the mode is selected, all of the “Function LEDs” will be OFF and the first error code
will be shown on the LED display. The user may scroll through the list by using the UP
and DOWN EDIT keys.
3.
Reset
This special mode selection allows the user to reset the unit (equivalent to a power-up
initialization).
This special mode is selected by pressing the “Select Function” key while simultaneously
pressing the UP EDIT key. The unit returns to the normal mode of operation (Monitoring
or Non-monitoring state) immediately after initialization.
4.
Store Setup in EEPROM
This special mode selection allows the user to force the unit to store all of the function
settings into non-volatile memory.
This special mode is selected by pressing the “Select Function” key while simultaneously
pressing the DOWN EDIT key. The unit returns to the normal mode of operation
(Monitoring or Non-monitoring state) immediately after writing all of the function settings
into non-volatile memory.
5.
Test Display
This special mode selection allows the user to test the display board by displaying a
different pattern on the LED indicators and the LED display every time any one of the
front panel buttons is pressed.
This special mode is selected by pressing the “Select Function” key while simultaneously
pressing the UP and LEFT EDIT keys. The unit returns to the normal mode of operation
(Monitoring or Non-monitoring state) after pressing the “Select Function” key one more time.
6.
Recall Default Setup
This special mode selection allows the user to recall the default function settings.
This special mode is selected by pressing the “Select Function” key while simultaneously
pressing the DOWN and RIGHT EDIT keys. The unit returns to the normal mode of
operation (Monitoring or Non-monitoring state) immediately after restoring all of the
default function settings.
Function
Factory Default Setting
1. Excitation Voltage
0.00 (OFF)
2. Sensor Sensitivity
1.000
3. Output Scaling
1.000
4. Low Pass Filter
XX.XX (ON)
5. Auto-Zero
OFF
6. Shunt Calibration
OFF
7. Monitoring State
Vout
7.
Calibration
This special mode selection allows the user to manually calibrate the unit.
This special mode is selected by pressing the “Select Function” key while simultaneously
pressing the LEFT and RIGHT EDIT keys. When this special mode is selected the first
“Function LED” will be blinking and the first calibration constant for the selected channel
will be shown on the LED display. The settings of all of the functions will be saved in non-
volatile memory when entering this mode.