IM7264B, Page 3
®
FIGURE 1 – Installation of Model 7264B Accelerometer
Recalibration
Sensitivity and ZMO calibration should be performed at 6 to 12 month intervals, depending on usage.
Ordinarily, recalibration need be performed only at 12 month intervals if it is known that the
accelerometer has not been used beyond its rated specifications. If the unit is used under severe
environments, it may be desirable to use shorter calibration intervals.
Contact Endevco for local calibration facility information or return the unit to Endevco for calibration.
Dirty units may be wiped clean using a damp cloth and a solvent such as acetone. DO NOT SOAK OR
IMMERSE the unit in any solvent or water. DO NOT USE any sharp tool such as a screwdriver to
remove dirt or contaminants.
Questions
If you have any questions regarding the use of this or any Endevco accelerometer, please contact
Endevco’s Application Engineering at 1-866-363-3826 in North America, or your local sales
representative.