![ENDEVCO 123 Instruction Manual Download Page 5](http://html1.mh-extra.com/html/endevco/123/123_instruction-manual_2406717005.webp)
5
CH1, CH2 and CH3 Signal Inputs
Reference the rear panel drawing shown on page 7.
Before connecting power to the instrument, verify INPUT POWER (ON/OFF) switch is in the OFF position
and the INPUT POWER SELECT switch is set to the correct voltage. Connect inputs and outputs prior
to applying power to the instrument.
For connection to PE sensors set INPUT SELECT switch to PE position and connect to the BNC input for
PE sensors.
For connection to IEPE sensors set INPUT SELECT switch to IEPE position and connect to the BNC
input for IEPE sensors. Select 4mA or 10mA source current for IEPE sensor excitation.
CH1, CH2 and CH3 Signal Outputs
Typical Hookup to an Oscilloscope