6.1 Flicker Load Operation
The typical flicker disturbance generated by the HFG01 in each of the steady flicker
modes when tested according to EN/IEC 61000-3-3 is shown in Table 3.
Flicker Mode
Setting
Load Operation
Pst
Plt
8.3 Hz
8.33 Hz flicker
1.10
1.10
1 Hz
1 Hz flicker
0.45
0.45
Table 3: Typical Pst results measured to EN/IEC 61000-3-3
6.2 Harmonics Load Operation
The typical current harmonics disturbance generated by the HFG01 in each of the
harmonics modes when tested according to EN/IEC 61000-3-2 is shown below:
Figure 7: Disturbances generated in Fixed Harmonic mode