8
Diagnostics and troubleshooting
8.1
pH/ORP diagnostics and troubleshooting
8.1.1
5081-P warning and fault messages
The 5081-P pH/ORP transmitter continuously monitors the measurement loop (sensor and
transmitter) for conditions that cause erroneous measurements. When a problem occurs,
the transmitter displays either a warning or fault message. A warning alerts you that a
potentially system disabling condition exists. If the condition causing the problem is not
corrected, there is a high probability that the system will soon fail. A fault alerts you that a
system disabling condition exists. If a fault message is showing, all measurements should
be regarded as erroneous.
When a Warning condition exists:
1.
The main display remains stable; it does not flash.
2.
A warning message appears alternately with the temperature display. See
.
for an explanation of the different warnings and suggested ways of
correcting the problem.
Warning annunciation
Figure 8-1:
When a Fault exists:
1.
The main display flashes.
2.
The words
FAULT
and
HOLD
appear in the main display.
3.
A fault message appears alternately with the temperature/output display. See
for an explanation of the different fault messages and
suggested ways of correcting the problem.
4.
The output current will remain at the present value or go to the programmed fault
value.
Diagnostics and troubleshooting
Instruction Manual
53
Summary of Contents for Rosemount5081
Page 4: ......
Page 10: ...Contents vi Rosemount 5081 ...
Page 12: ...Startup procedure 2 Rosemount 5081 ...
Page 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Page 24: ...Installation 14 Rosemount 5081 ...
Page 28: ...Wiring 18 Rosemount 5081 ...
Page 50: ...Programming basics 40 Rosemount 5081 ...
Page 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Page 136: ...Digital communications 126 Rosemount 5081 ...
Page 140: ...Engineering drawings 130 Rosemount 5081 ...
Page 141: ...Engineering drawings Instruction Manual 131 ...
Page 142: ...Engineering drawings 132 Rosemount 5081 ...
Page 143: ...Engineering drawings Instruction Manual 133 ...
Page 144: ...Engineering drawings 134 Rosemount 5081 ...
Page 145: ...Engineering drawings Instruction Manual 135 ...
Page 146: ...Engineering drawings 136 Rosemount 5081 ...
Page 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...