11
Reference Manual
00809-0100-4021, Rev GC
Section 2: Installation
April 2014
2.2.2
Set the switches
HART
®
Without an LCD display
1.
If the transmitter is installed in a loop, set the loop to manual mode and disconnect the
power.
2.
Remove the housing cover on the electronics side of the transmitter. Do not remove
the transmitter cover in explosive atmospheres with a live circuit.
3.
Set the switches to the desired position (see
).
4.
Replace the transmitter cover. Both transmitter covers must be fully engaged to meet
explosion-proof requirements.
5.
Apply power and set the loop to automatic mode.
With an LCD display
1.
If the transmitter is installed in a loop, set the loop to manual mode and disconnect the
power.
2.
Remove the housing cover on the electronics side of the transmitter. Do not remove
the transmitter cover in explosive atmospheres with a live circuit.
3.
Unscrew the LCD display screws and gently slide the meter straight off.
4.
Set the switches to the desired position (see
).
5.
Gently slide the LCD display back into place, taking extra precautions with the 10 pin
connection.
6.
Replace and tighten the LCD display screws to secure the LCD display.
7.
Replace the transmitter cover. Both transmitter covers must be fully engaged to meet
explosion-proof requirements.
8.
Apply power and set the loop to automatic mode.
Summary of Contents for Rosemount
Page 1: ...Reference Manual 00809 0100 4021 Rev GC April 2014 Rosemount 3144P Temperature Transmitter ...
Page 2: ......
Page 4: ...iv Reference Manual 00809 0100 4021 Rev GC Title Page April 2014 ...
Page 10: ...x Reference Manual 00809 0100 4021 Rev GC Table of Contents April 2014 Table of Contents ...
Page 38: ...28 Reference Manual 00809 0100 4021 Rev GC Section 2 Installation April 2014 Installation ...
Page 188: ...178 Reference Manual 00809 0100 4021 Rev GC Section 5 Maintenance April 2014 Maintenance ...
Page 250: ...Index 244 Reference Manual 00809 0100 4021 Rev GC Index April 2014 Index ...
Page 251: ......