32
Reference Manual
00809-0200-4648, Rev BB
July 2016
Figure 3-5. Direct Mount Installation
3.5.2
Remote mount
1.
Install the sensor/thermocouple according to standard installation practices. Be
sure to use thread sealant on all connections.
2.
Run wiring (and conduit if necessary) from the sensor/thermocouple to the
Rosemount 648 Wireless.
3.
Pull the wiring through the threaded conduit entry of the Rosemount 648 Wireless.
4.
Attach the sensor/thermocouple wiring to the terminals as indicated on the wiring
diagram.
5.
Connect the power module if commissioning.
Note
Use caution when handling the power module. The power module it may be damaged if
dropped from heights in excess of 20 feet.
Possible antenna rotation shown.
Antenna rotation allows for best
installation practices in any
configuration.