X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
7
Maintenance & Procedures
7.4.3 Advanced Calibration
7.4.3.4 Calibrating With Sequence Programming
Setup..
Calibration..
Program Sequence..
...
Program Sequence 4of8
Action13:
END-OF-PGRM
Node13:
All
Action14:
END-OF-PGRM
Node14:
All
Action15:
END-OF-PGRM
Node15:
All
Action16:
END-OF-PGRM
Node16:
All
Program Sequence 8of8
Action29:
END-OF-PGRM
Node29:
All
Action30:
END-OF-PGRM
Node30:
All
Page 1
Page 4
Page 8
Program Sequence 1of8
Action1:
Zero-Cal
Node1:
All
Action2:
Rg1SpanCal
Node2:
Ch1
Action3:
Rg2SpanCal
Node3:
Ch3
Action4:
END-OF-PGRM
Node4:
All
This menu with 8 pages allows to setup a
sequences of up to 30 actions (steps), to
carry out individual calibration procedures.
Each step consists of an action and a related
node (channel).
Available actions are:
Action name
What to happen
Rg1SpanCal ...
Rg4SpanCal
span calibrate range1
... range4
ZSpanCal
zero & span calibrate
SpanCal
span calibrate
ZeroCal
zero calibrate
NoOp
no action
Blowback
start blowback
END-OF-PGRM
end of programmed
sequence
Available nodes are (depending on number
of channels installed within your analyzer):
Node name
Selected action is
carried out for
All
all installed channels
Ch1 ... Ch5
(depending on the
analyzer setup, an
assigned component
tag may show here
instead;
the selected channel
only
Program sequences enable to carry out
complex calibration procedures with up to
30 steps. A requirement to setup a sequence
is, that “Valves” in I
NSTALLED OPTIONS
is
set to a value other than
none
(
Once this condition is fulfilled, a sequence
can be setup in SETUP - CALIBRATION -
PROGRAM SQEUENCE.