Procedure
1.
Unplug the connector from the board and wire the power cable to it.
Lead connections are marked on the connector. (
L
is live or hot;
N
is neutral; the
ground connection has the standard symbol.)
2.
AC power wiring should be 14 gauge or greater. Run the power wiring through the
conduit opening nearest the power terminal.
3.
Provide a switch or breaker to disconnect the transmitter from the main power
supply.
4.
Install the switch or breaker near the transmitter and label it as the disconnecting
device for the transmitter.
4.3.2
Analog output wiring
Two analog current outputs are located on the main circuit board, which is attached to the
inside of the enclosure door.
shows the location of the terminals. The connectors can be detached for wiring.
TB-1 is output 1. TB-2 is output 2. Polarity is marked on the circuit board.
Analog output connections
Figure 4-2:
The analog outputs are on the main board near the hinged end of the enclosure door.
Wiring
Instruction Manual
21
Summary of Contents for Rosemount TCL
Page 4: ......
Page 8: ...Contents iv Rosemount TCL ...
Page 16: ...Description and specifications 8 Rosemount TCL ...
Page 18: ...Principles of operation 10 Rosemount TCL ...
Page 26: ...Installation 18 Rosemount TCL ...
Page 34: ...Wiring 26 Rosemount TCL ...
Page 46: ...Display and operation 38 Rosemount TCL ...
Page 66: ...Programming the transmitter 58 Rosemount TCL ...
Page 68: ...Digital communications 60 Rosemount TCL ...
Page 106: ...Troubleshooting 98 Rosemount TCL ...