Quick Start Guide
9
February 2018
Markings: Intrinsically Safe for Class I, Division 1, Groups A, B, C, D T4;
Ex ia IIC T4 Ga
T4 (–50 °C
Ta
+70 °C)
when installed per Rosemount drawing 00G45-1020;
Type 4X; IP66/67;
Special Conditions for Safe Use (X):
1. Do not replace battery when explosive atmosphere is present.
Ne pas remplacer les accumulateurs si une atmosphère explosive peut être
présente.
2. Use only 00G45-9000-0001 batteries.
Utiliser uniquement des accumulateurs 00G45-9000-0001.
3. The surface resistivity of the housing is greater than 1G
. To avoid electrostatic
charge build-up, it must not be rubbed or cleaned with solvents or a dry cloth.
La résistivité de surface du boÎtier est supérieure à un gigaohm. Pour éviter
l’accumulation de charge électrostatique, ne pas frotter ou nettoyer avec des
produits solvants ou un chiffon sec.
4. Substitution of components may impair intrinsic safety.
La substitution de composants peut compromettre la sécurité intrinsèque.
5.6
Europe
I1
ATEX Intrinsic Safety
Certificate: Baseefa16ATEX0005X
Standards: EN 60079-0: 2012 + A11: 2013, EN 60079-11: 2012
Markings:
II 1 G Ex ia IIC T4 Ga, T4 (–40 °C
Ta
+70 °C)
IP66/67;
Special Conditions for Safe Use (X):
1. The plastic enclosure may constitute a potential electrostatic ignition risk and must
not be rubbed or cleaned with a dry cloth.
2. The measured capacitance between the equipment enclosure and metallic inline
sensor module is 4.7pF. This must be considered only when the SPG is integrated
into a system where the process connection is not grounded.
3. Do not change the battery when an explosive atmosphere is present.
4. Only replace battery with Rosemount Part No. 00G45-9000-0001.
5.7
International
I7
IECEx Intrinsic Safety
Certificate: IECEx BAS 16.0012X
Standards: IEC 60079-0: 2011, IEC 60079-11: 2011
Markings: Ex ia IIC T4 Ga, T4 (–40 °C
T
a
+70 °C)
IP66/67;
Special Conditions for Safe Use (X):
1. The plastic enclosure may constitute a potential electrostatic ignition risk and must
not be rubbed or cleaned with a dry cloth.
2. The measured capacitance between the equipment enclosure and metallic inline
sensor module is 4.7pF. This must be considered only when the SPG is integrated
into a system where the process connection is not grounded.
3. Do not change the battery when an explosive atmosphere is present.
4. Only replace battery with Rosemount Part No. 00G45-9000-0001.