B
Upgrade Rosemount
™
Oxymitter DR
to full Oxymitter
Figure B-1: Component Replacement
A. Screws
B. Rosemount OXT adapter board
C. Screws
D. Electronic assembly
Procedure
1. Remove power from the Rosemount Oxymitter DR.
2. Remove the left and right covers from the Rosemount Oxymitter termination
housing.
3. Remove and discard screws and Rosemount OXT adapter board (A and B,
) located at the right side of the termination housing.
4. Place the new Rosemount Oxymitter electronic assembly (D) near the right side of
the termination housing.
5. Plug the white connector with the two black wires into the white socket on the
bottom power supply card.
6. Insert the electronics assembly into the termination housing. Ensure the black four-
wire connector remains outside the housing and in the slot provided in the top card
of the electronics assembly.
The electronics assembly should seat on the bulkhead pins easily. Do not force the
assembly into place.
Reference Manual
Upgrade Rosemount
™
Oxymitter DR to full Oxymitter
00809-0100-4880
October 2019
Rosemount Oxymitter DR
87