Figure 4-4: Ceramic Diffuser Element Replacement
A. Threaded probe flange
B. M5-0.8 x 5 mm locking set screw (use 2.5 mm hex key)
C. Spanner wrench
D. Ceramic diffusion element
E. M5-1 x 6 mm setscrew (use 3 mm hex key)
F. Vee deflector
G. Cement fillet
H. Cement port
I. Flame arrestor hub
4. If equipped with the optional ceramic diffusion assembly, remove and discard the
setscrews and remove the vee deflector (
) to turn the hub free from the retainer. Inspect
the diffusion element. If damaged, replace the element.
5. Loosen the four socket head cap screws from the cell and flange assembly.
6. Remove the assembly and the corrugated cell.
Reference Manual
Maintenance and service
00809-0100-4880
October 2019
Rosemount Oxymitter DR
61