Figure 4-2: Heater Strut Assembly
A. V-deflector
B. Ceramic support rod
C. Wire loop
D. Ceramic diffuser assembly
E. Cell flange
F. Heater
8. When replacing the strut, orient the probe so that the small calibration gas tuble
lays at the 6 o'clock position of the probe tube. Align the slot on the heater plate
with the calibration gas line in the probe tube. Slide the strut into the probe tube.
The strut turns to align the hole on the back plate of the strut with the calibration
gas line. When the hole and the calibration gas line are aligned correctly, the strugt
will slide in the rest of the way.
9. As the strut installation nears completion, install the guide rod into the calibration
gas tube to assist in guiding the calibration gas tube through the hole in the end of
the strut.
10. Push down the back plate of the strut to make sure you have spring tension and
then tighten the three screws on the back plate.
11. Replace the CAL and REF gas silicone tubes.
12. Install the termination housing per
.
13. Follow the instructions in
to install the probe in the stack or duct.
Maintenance and service
Reference Manual
October 2019
00809-0100-4880
58
Emerson.com/Rosemount