Do not get any cement on upper part of ceramic diffusion element. Ensure
complete penetration of cement around three grooves in hub. Cement should
extrude from opposite hole.
10. Wipe excess material back into holes and wipe top fillet of cement to form a
uniform fillet.
A cotton swab is useful for this.
11. Clean any excess cement from hub with water.
12. Allow filter to dry at room temperature overnight or one to two hours at 200 °F (93
°C).
13. Wipe a heavy layer of anti-seize compound onto the threads and mating surfaces of
the flame arrestor, diffusion hub, and probe tube.
14. On systems equipped with abrasive shield, install dust seal gaskets with joints 180 °
apart.
15. Reinstall vee deflector, orienting apex towards gas flow.
16. Apply anti-seize compound to set-screws and tighten with hex wrench.
17. Reinstall probe on stack flange.
9.3.11
Replace contact and thermocouple assembly
Figure 9-9: Contact and Thermocouple Assembly Replacement
A. Strut
B. Contact and thermocouple assembly
C. Spring
D. Spring clip
E. Tube support
Procedure
.
2. Remove the heater strut assembly, per
, steps
through
.
Reference Manual
Maintenance and service
00809-0100-4340
October 2019
Rosemount Oxymitter 4000
171
Summary of Contents for Rosemount Oxymitter 4000
Page 4: ...4 ...
Page 8: ...Contents Reference Manual October 2019 00809 0100 4340 viii Emerson com Rosemount ...
Page 110: ...HART AMS Reference Manual October 2019 00809 0100 4340 110 Emerson com Rosemount ...
Page 114: ...HART AMS Reference Manual October 2019 00809 0100 4340 114 Emerson com Rosemount ...
Page 184: ...Optional accessories Reference Manual October 2019 00809 0100 4340 184 Emerson com Rosemount ...