Section 3: Configuration and
Startup
April 2017
Instruction Manual
IM-106-880C, OI
50
Performing a Calibration Verify with the Field Communicator -
HART
The calibration verify feature is only valid if the Hazardous Area OCX 8800 is supplied with
calibration solenoids and the solenoids have been activated.
During the Calibration Verify function the analog output signals will track the oxygen and
combustibles readings. To avoid a potentially dangerous operating condition, the
Hazardous Area OCX 8800 must be removed from the automatic combustion control
loop before performing the Calibration Verify procedure.
3.7.1
Performing a Calibration Verify with the Field
Communicator - HART
1.
Use the 375 Field Communicator or AMS software to access the HART menu.
2.
From the DEVICE SETUP menu, select DIAG/SERVICE.
3.
From the DIAG/SERVICE menu, select CALIBRATION.
4.
From the CALIBRATION, select CAL VERIFY. Select Verify Calibration. From this menu,
select the functions as follows:
•
Flow High O2 Gas - Flows the high O2 test gas for the time specified in the
calibration setup.
•
Flow Low O2 Gas - Flows the low O2 test gas for the time specified in the
calibration setup.
•
Flow High COe Gas - Flows the COe test gas for the time specified in the
calibration setup.
•
Purge Gas - Initiates a delay for the specified purge time before oxygen/
combustibles readings are considered valid.
NOTE:
A Purge will automatically follow a gas flow.
3.7.2
Performing a Calibration Verify with the Field
Communicator - Fieldbus
1.
Use the 375 Field Communicator or AMS software to access the Fieldbus menu.
2.
From the TRANSDUCER block menu, select METHODS.
3.
Set the Mode to OOS (Out of Service) before starting the Calibration Verify process.
Summary of Contents for Rosemount OCX 8800
Page 2: ......
Page 41: ...Instruction Manual IM 106 880C OI Section 2 Installation April 2017 Section contents 17 ...
Page 205: ...Alarm Relay Events April 2017 Instruction Manual IM 106 880C OI 17 8 Alarm Relay Events ...
Page 251: ...Replace Tube Fittings April 2017 Instruction Manual IM 106 880C OI 22 4 Install Tube Fittings ...
Page 261: ...Instruction Manual IM 106 880C OI Safety Data April 2017 Appendix contents 233 ...
Page 263: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 235 ...
Page 267: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 239 ...
Page 271: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 243 ...
Page 273: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 245 ...
Page 275: ...Instruction Manual Safety Data IM 106 880C OI April 2017 IMPORTANT Safety Instructions 247 ...
Page 277: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 249 ...
Page 279: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 251 ...
Page 300: ...Instruction Manual IM 106 880C OI Index April 2017 Manual index 27 4 ...
Page 304: ...Index April 2017 Instruction Manual IM 106 880C OI 27 8 Manual index ...
Page 305: ......