13
Preventative maintenance
13.1
Maintaining the analyzer
WARNING
MAINTENANCE / MODIFICATIONS
On completion of any maintenance and/or modifications, verify:
—
All tools and equipment are removed.
—
No contamination (water or dust) is in the compartments.
—
The analyzer is wiped clean.
—
Vents are clear and not obstructed.
—
The system is in a safe state for operation.
13.2
Scheduled maintenance
This schedule lists the tasks required by the analyzer and the recommended frequency.
Variation in customer sites may require these activities to be performed more or less often
than indicated. Details of the tasks to be performed are contained in
.
Table 13-1: Scheduled Checks
Frequency
Action
Monthly
1. Check the zero and span calibration. Perform the calibration more or
less frequently if necessary to meet quality control or plant
operation requirements.
2. Clean outside of system. Ensure all vents are clean and
unobstructed.
Every three months
(quarterly)
1. Perform the monthly check detailed in the previous row.
2. Purge the long path cell mirrors.
3. Check the X-Shift calibration of the lasers. Contact your local service
representative for guidance.
Annually
1. Perform the three-monthly check detailed in the previous row.
2. Perform a laser wavelength calibration. Contact your local service
representative for guidance.
Reference Manual
Preventative maintenance
00809-0100-4512
October 2019
Reference Manual
163
Summary of Contents for Rosemount CT5100
Page 16: ...16 ...
Page 20: ...Contents Reference Manual October 2019 00809 0100 4512 xx Rosemount CT5100 Ex ...
Page 28: ...Theory of operation Reference Manual October 2019 00809 0100 4512 28 Rosemount CT5100 Ex ...
Page 46: ...Specifications Reference Manual October 2019 00809 0100 4512 46 Rosemount CT5100 Ex ...
Page 66: ...Install Reference Manual October 2019 00809 0100 4512 66 Rosemount CT5100 Ex ...
Page 84: ...Start up procedure Reference Manual October 2019 00809 0100 4512 84 Rosemount CT5100 Ex ...
Page 92: ...Shutdown procedure Reference Manual October 2019 00809 0100 4512 92 Rosemount CT5100 Ex ...
Page 112: ...Gas calibration Reference Manual October 2019 00809 0100 4512 112 Rosemount CT5100 Ex ...