15. Secure the gas temperature sensor wiring harness with cable ties and cable clips in
the locations that were noted down in
.
16. Fit the top cover and restart the analyzer.
12.20
Replacing the pressure sensor
Figure 12-17: Replacing the Pressure Sensor
A. Bulkhead adaptor
B. Dowty washer
C. Pressure sensor
D. Pressure sensor electrical connector
Procedure
1. Shut down the analyzer.
.
2. Remove the analyzer top cover.
.
3. Disconnect the pressure sensor electrical connector (D).
4. While holding the bulkhead adaptor (A) steady unscrew the pressure sensor (C).
5. Discard the unserviceable pressure sensor.
6. Inspect the new pressure sensor for damage.
7. Install the replacement pressure sensor (C) and dowty washer (B) onto the bulkhead
adaptor (A). Hold the bulkhead adaptor steady while tightening the pressure
sensor.
8. Reconnect the pressure sensor electrical connector (D) to the pressure sensor.
Reference Manual
Repair and replacement
00809-0100-4512
October 2019
Reference Manual
155
Summary of Contents for Rosemount CT5100
Page 16: ...16 ...
Page 20: ...Contents Reference Manual October 2019 00809 0100 4512 xx Rosemount CT5100 Ex ...
Page 28: ...Theory of operation Reference Manual October 2019 00809 0100 4512 28 Rosemount CT5100 Ex ...
Page 46: ...Specifications Reference Manual October 2019 00809 0100 4512 46 Rosemount CT5100 Ex ...
Page 66: ...Install Reference Manual October 2019 00809 0100 4512 66 Rosemount CT5100 Ex ...
Page 84: ...Start up procedure Reference Manual October 2019 00809 0100 4512 84 Rosemount CT5100 Ex ...
Page 92: ...Shutdown procedure Reference Manual October 2019 00809 0100 4512 92 Rosemount CT5100 Ex ...
Page 112: ...Gas calibration Reference Manual October 2019 00809 0100 4512 112 Rosemount CT5100 Ex ...