11
Troubleshooting and diagnostics
11.1
Using the Built-in-Self-Test (BIST) fault
diagnostics
The analyzer has a Built-In-Self-Test (BIST) function that can be used to perform failure
diagnosis of some functions in situations where there is a fault other than a complete
failure of the equipment. The BIST is accessed and controlled through the display
controller mounted on the door of the electrical compartment.
The BIST runs in the background when the analyzer is operating and continuously
monitors the analyzer for faults.
Procedure
1. If the BIST detects a fault, the
STATUS
display in the bottom left-hand corner of the
Gas Sensor Main
OK
to flashing the
word
FAILURE
.
Figure 11-1: Gas Sensor Main Screen
A. FAILURE message
B. FAILURE button
2. Similarly, if a fault is detected and the
Pressure and Temperature
)
is being displayed, the
STATUS
display in the bottom left-hand corner of the screen
(A) will change from
OK
to flashing the word
FAILURE
.
3. On either the
Gas Sensor Main
Pressure and Temperature
screen (
), pressing the
FAILURE
button (B) will take you to the
Faults
menu
).
Reference Manual
Troubleshooting and diagnostics
00809-0100-4512
October 2019
Reference Manual
113
Summary of Contents for Rosemount CT5100
Page 16: ...16 ...
Page 20: ...Contents Reference Manual October 2019 00809 0100 4512 xx Rosemount CT5100 Ex ...
Page 28: ...Theory of operation Reference Manual October 2019 00809 0100 4512 28 Rosemount CT5100 Ex ...
Page 46: ...Specifications Reference Manual October 2019 00809 0100 4512 46 Rosemount CT5100 Ex ...
Page 66: ...Install Reference Manual October 2019 00809 0100 4512 66 Rosemount CT5100 Ex ...
Page 84: ...Start up procedure Reference Manual October 2019 00809 0100 4512 84 Rosemount CT5100 Ex ...
Page 92: ...Shutdown procedure Reference Manual October 2019 00809 0100 4512 92 Rosemount CT5100 Ex ...
Page 112: ...Gas calibration Reference Manual October 2019 00809 0100 4512 112 Rosemount CT5100 Ex ...