CALIBR PARAMS
allows configuration of the purge time for Gas Calibration.
VERIFY PARAMS
allows configuration of the purge time for Gas Verification.
STREAM SWITCH
not an option available for the Rosemount CT4400 analyzer.
Pressing the
Back
key
will return to the previous screen. Pressing the
Home
key
at
any time will return to the Main Menu screen.
6.3
Calibration
Calibration flows the known cylinder gas through the analyzer and then adjusts the
readout until the measurement matches the cylinder. The analyzer then applies this
adjustment to all measurements until the next calibration. There are two circumstances in
which you may want to calibrate:
1. After you verify gas concentrations (see
) and find that
the analyzer readings are outside of tolerance.
2. At regular intervals, such as once a month.
6.3.1
Zero calibration
The zero calibration procedure adjusts for any zero offset in the measurement. Zero
calibration is done by calibrating the analyzer measurements against a known sample gas
using the following procedure.
Prerequisites
Run the analyzer at a stable temperature for at least thirty minutes prior to commencing
this procedure.
Procedure
1. Use nitrogen gas of "instrument gas" purity as the zero calibration gas.
2. Ensure that a pressure regulator is connected to the nitrogen gas bottle.
3. Connect a hose from the nitrogen gas bottle through a T-piece to the sample supply
port on the rear of the analyzer.
4. Connect an excess flow line to the unused port on the T-piece and route the excess
flow line to a suitable extractor.
5. On the display controller of the analyzer, browse to the
Main menu
as described in
Gas calibration
Reference Manual
January 2020
00809-0100-4244
50
Rosemount CT4400
Summary of Contents for Rosemount CT4400
Page 1: ...Reference Manual 00809 0100 4244 Rev AA January 2020 Rosemount CT4400 Continuous Gas Analyzer ...
Page 32: ...Start up procedure Reference Manual January 2020 00809 0100 4244 32 Rosemount CT4400 ...
Page 40: ...Operating the analyzer Reference Manual January 2020 00809 0100 4244 40 Rosemount CT4400 ...
Page 64: ...Gas calibration Reference Manual January 2020 00809 0100 4244 64 Rosemount CT4400 ...
Page 96: ...Shutdown procedure Reference Manual January 2020 00809 0100 4244 96 Rosemount CT4400 ...
Page 98: ...Preventative maintenance Reference Manual January 2020 00809 0100 4244 98 Rosemount CT4400 ...
Page 100: ...Recommended spare parts Reference Manual January 2020 00809 0100 4244 100 Rosemount CT4400 ...
Page 105: ......