Set the calibration target either to a known value of the gas concentration or to zero. Then
set the desired number of cycles over which the calibration factor is determined
(Rosemount
™
recommends a minimum of 30). Now run the calibration, and the display will
show a countdown during its execution. When the calibration is complete, the new
Cal
Factor
is displayed for about five seconds, and the instrument exits the calibration routine.
Note
You must run the calibration routine during commissioning; otherwise, the instrument will not be
able to calculate the true level of gas within the duct.
Do not run the calibration routine unless reasonable conditions exist in the duct. If it is not
the initial calibration, record the
Cal Factor
from the
Parameters
option before running the
calibration.
To aid configuration and to record any subsequent changes to the operating parameters,
lists all of the options available and can be used as a record of the operating
parameters.
Instrument Settings
Table 3-1:
Parameter
Config
Averages
Seconds
Minutes
Hours
Days
Output
0 or 4 mA base
Units
Average
Configuration and startup
48
Rosemount CCO 5500
Summary of Contents for Rosemount CCO 5500
Page 4: ......
Page 8: ...Contents iv Rosemount CCO 5500 ...
Page 28: ...System Wiring Diagram Figure 2 7 Install 20 Rosemount CCO 5500 ...
Page 32: ...Install 24 Rosemount CCO 5500 ...
Page 88: ...Operation 80 Rosemount CCO 5500 ...
Page 98: ...Maintenance 90 Rosemount CCO 5500 ...
Page 108: ...Troubleshooting 100 Rosemount CCO 5500 ...
Page 110: ...Spare parts 102 Rosemount CCO 5500 ...
Page 122: ...Return equipment to the factory 114 Rosemount CCO 5500 ...
Page 123: ...Return equipment to the factory Reference Manual 115 ...