Procedure
1.
Press MODE unil the number
5
is displayed in the top left corner.
After you correctly enter the security code, there are six submodes of operation
from which the set up parameters may be changed. These six submodes are:
a. Set Averages: The four averaging stack times (seconds, minutes, hours, and
days) may be set as required.
b. Configure O/P - Analog output setup: Origin, units, span, rolling average, and
fault condition.
c. Parameters: The following are set from this mode: security code, identity
number, path length, alarm level, cal factor, and plant status.
d. Normalization: You may set up all normalization parameters from this mode.
e. Reset Average: Select this submode to reset the four averaging stacks.
f. Calibrate: Set the outputs of the detectors and the basic calibration for the
instrument.
2.
Use the arrow keys to toggle between these six options and press ENTER when the
desired option is displayed.
3.7.1
Enter security code
Complete the following steps to enter the security code.
1.
Once the display is as shown here, press ENTER to gain access to
SET UP
mode.
The cursor flashes over the first digit of the security code number.
2.
Select the required first digit with the arrow keys and press ENTER.
3.
Repeat this procedure for the remaining three numbers.
If the code is correct after you press ENTER on the last digit, the sequence is
continued. If it is not correct, the instrument returns to
OPERATING
mode. Refer to
Note
Rosemount
™
sets the code number to
0000
at the factory; you should change it in
SET UP
mode.
3.7.2
Set averages
The instrument calculates four separate averages. These are defined in units of seconds,
minutes, hours, and days. You can use any of the four averaging stacks to provide the
instrument's analog output. You can set each averaging time within predefined limits.
Configuration and startup
36
Rosemount CCO 5500
Summary of Contents for Rosemount CCO 5500
Page 4: ......
Page 8: ...Contents iv Rosemount CCO 5500 ...
Page 28: ...System Wiring Diagram Figure 2 7 Install 20 Rosemount CCO 5500 ...
Page 32: ...Install 24 Rosemount CCO 5500 ...
Page 88: ...Operation 80 Rosemount CCO 5500 ...
Page 98: ...Maintenance 90 Rosemount CCO 5500 ...
Page 108: ...Troubleshooting 100 Rosemount CCO 5500 ...
Page 110: ...Spare parts 102 Rosemount CCO 5500 ...
Page 122: ...Return equipment to the factory 114 Rosemount CCO 5500 ...
Page 123: ...Return equipment to the factory Reference Manual 115 ...