1.
. Unscrew the four screws that secure the receiver to the air purge.
Remove and place the receiver in a safe location.
2.
Go to the source unit location. To align the source, turn the adjusting nuts
(
). Use opposing adjusting nuts to align the source unit in one plane, then
the other.
Note
You can achieve receiver alignment by monitoring the output of the detector directly. Use a
voltmeter set to AC volts (10 V max.) to measure across test points S0V and S2 for D3, and
S0V and S1 for D1 on the receiver control board (
). This alignment method is useful
when the receiver is not located near the control unit.
Alignment Features
Figure 3-2:
A. Adjusting nuts
B. Locking nuts
C. Air purge
D. Rear flange
E. Front flange
Configuration and startup
Reference Manual
27