Instruction Manual
IM-105-4000, Rev 1.3
October 2008
OPM 4000
http://www.raihome.com
Appendix C
ModBus Communications
MODBUS
COMMUNICATIONS
The OPM 4000 is capable of Modbus communications. To communicate with
the OPM 4000 via Modbus, connect the Modbus signal wires to the Isolated
RS 485 terminals 1, 2, and 3 on the back of the Control Room Unit. These
terminals are shown in Figure 2-11. The Modbus addresses used by the OPM
4000 are shown in Table C-1.
Table C-1. Modbus Addresses
Note 1: Air purge low, No stack power, T2 4-20 mA lost/low, Maintenance mode, Window dust, Zero cal
fail, Span cal fail, and Manual cal
Read Block
Address
Function
40150
0% to 100% instantaneous opacity output (analog 4-20 mA)
40151
0% to 100% average opacity output (analog 4-20 mA)
00100
Cal zero check
00101
Cal span check
00140
Cal zero fail alarm
00141
Cal span fail alarm
00042
Loss of 120 VAC transceiver power
00075
Dust/Zero shift alarm
00097
High opacity alarm
00098
Common fault
(Note 1)
Write Block
Address
Function
45
Remote Cal Initiate
Summary of Contents for Rosemount Analytical OPM 4000
Page 2: ......
Page 6: ......
Page 10: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 TOC 4 ...
Page 14: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 iv ...
Page 42: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 2 20 ...
Page 52: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 4 6 ...
Page 58: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 5 6 ...
Page 64: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 6 6 ...
Page 100: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 A 24 ...
Page 102: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 B 2 ...
Page 104: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 C 2 ...
Page 114: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 D 10 ...
Page 116: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 Index 2 OPM 4000 ...