6.3
Digital concentration output is not reacting to
presence of target gas
Potential cause
Clogged IP filter
Recommended action
Check whether the IP filter is clogged; replace if necessary.
Refer to
Replace the Ingress Protection filter
Potential cause
Gas sensor module is at the end of its service life.
Recommended action
Replace with a new module.
6.4
LCD display troubleshooting
6.4.1
LCD display is not operating
Potential cause
LCD display is not enabled.
Recommended action
Make sure the LCD display is enabled.
Potential cause
Connector is missing or invalid.
Recommended action
Make sure the LCD display pins are present and not bent.
Potential cause
LCD display is not fully connected.
Recommended action
Make sure the LCD display is properly seated with the tabs snapped in place and fully
engaged.