Continuous SMART Meter Verification Test Parameters
(continued)
Table 9-4:
Parameter
LOI menu path ( Diagnostics > Varia-
bles > MV Results > Continual Res )
8
Electrode Resistance
Electrode Res
9
mA Expected
4–20 mA Expect
10
mA Actual
4–20 mA Actual
11
mA Deviation
AO FB Dev
9.13
SMART Meter Verification measurements
The SMART Meter Verification test will make measurements of the coil resistance, coil
signature, and electrode resistance and compare these values to the values taken during
the sensor signature process to determine the sensor calibration deviation, the coil circuit
health, and the electrode circuit health. In addition, the measurements taken by this test
can provide additional information when troubleshooting the meter.
Coil circuit resistance
LOI menu path
Manual: Diagnostics > Advanced Diag > Meter Verif > Measure-
ments > Manual Measure > Coil Resist
Continuous: Diagnostics > Advanced Diag > Meter Verif > Meas-
urements > Continual Meas > Coil Resist
The coil circuit resistance is a measurement of the coil circuit health. This value is
compared to the coil circuit resistance baseline measurement taken during the sensor
signature process to determine coil circuit health. This value can be continuously
monitored using continuous SMART Meter Verification.
Coil signature
LOI menu path
Manual: Diagnostics > Advanced Diag > Meter Verif > Measure-
ments > Manual Measure > Coil Inductnce
Continuous: Diagnostics > Advanced Diag > Meter Verif > Meas-
urements > Continual Meas > Coil Inductnce
The coil signature is a measurement of the magnetic field strength. This value is compared
to the coil signature baseline measurement taken during the sensor signature process to
determine sensor calibration deviation. This value can be continuously monitored using
continuous SMART Meter Verification.
Advanced Diagnostics Configuration
Reference manual
127
Summary of Contents for Rosemount 8732EM
Page 2: ......
Page 6: ...Contents iv Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 10: ...Safety messages 4 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 14: ...Introduction 8 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 30: ...Sensor Installation 24 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 92: ...Operation 86 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 158: ...Maintenance 152 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 214: ...Product Certifications 208 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 217: ...Wiring 8732EM using combination cable Figure C 2 Wiring Diagrams Reference manual 211 ...
Page 222: ...Wiring Diagrams 216 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 253: ...Implementing a Universal Transmitter Reference manual 247 ...