3
Consider housing rotation
The entire electronics housing may be rotated in 90° increments for easy
viewing. Use the following steps to change the housing orientation,
1. Loosen the four housing rotation set screws at the base of the electronics
housing with a 5/32” hex wrench by turning the screws clockwise
(inward) until they clear the support tube.
2. Slowly pull the electronics housing out of the support tube.
CAUTION!
Do not pull the housing more than 1.5 in. (40 mm) from the top of the
support tube until the sensor cable is disconnected. Damage to the
sensor may occur if this sensor cable is stressed.
3. Unscrew the sensor cable from the housing with a 5/16” open end
wrench.
4. Rotate the housing to the desired orientation.
5. Hold it in this orientation while you screw the sensor cable onto the base
of the housing.
CAUTION!
Do not rotate the housing while the sensor cable is attached to the base
of the housing. This will stress the cable and may damage the sensor.
6. Place the electronics housing into the top of the support tube.
7. Use a hex wrench to turn the four housing rotation screws counter-
clockwise (outward) to engage the support tube.
April 2018
Quick Start Guide
Quick Start Guide
11
Summary of Contents for Rosemount 8600D Series
Page 1: ...Quick Start Guide 00825 0100 4860 Rev EB April 2018 Rosemount 8600 Series Vortex Flowmeter ...
Page 35: ...April 2018 Quick Start Guide Quick Start Guide 35 ...
Page 36: ...Quick Start Guide April 2018 36 Rosemount 8600 Series Vortex Flowmeter ...
Page 37: ...April 2018 Quick Start Guide Quick Start Guide 37 ...
Page 38: ...Quick Start Guide April 2018 38 Rosemount 8600 Series Vortex Flowmeter ...
Page 39: ...April 2018 Quick Start Guide Quick Start Guide 39 ...