IECEx Intrinsic Safety for Zone 2
Certificate: IECEx BAS 14.0173X
Standards: IEC 60079-0: 2011, IEC 60079-11: 2011
Markings: Ex ic IIC T4 Gc, T4 (-60 °C
≤
T
a
≤
+70 °C)
For use with Rosemount SmartPower power module part
number 753-9220-0001, or for use with Emerson
SmartPower option 701PBKKF.
See the table at the end of this section for entity parameters
Special Conditions for Safe Use (X):
1. The surface resistivity of the antenna is greater than
1 G
Ω
. To avoid electrostatic charge build-up, it must not be
rubbed or cleaned with solvents or a dry cloth.
2. The Rosemount 705 Transmitter enclosure may be made of
aluminum alloy and given a protective polyurethane paint
finish; however, care should be taken to protect it from
impact or abrasion if located in a Zone 0 area.
B.10 Brazil
I2
INMETRO Intrinsic Safety
Certificate: UL-BR 17.0019X
Standards: ABNT NBR IEC 60079-0:2008 + Errata 1:2011,
ABNT NBR IEC 60079-11:2009
Markings: Ex ia IIC T4 Ga, T4 (–60 °C
≤
T
a
≤
+70 °C)
See the table at the end of this section for entity parameters
Special Condition for Safe Use (X):
1. See certificate for special conditions.
B.11 EAC - Belarus, Kazakhstan, Russia
IM
Technical Regulation Customs Union (EAC) Intrinsic Safety
Certificate: TC RU C-US.MIO62.B.03122
Markings: 0Ex ia IIC T4 Ga X, T4 (–60 °C
≤
T
a
≤
+70 °C)
See the table at the end of this section for entity para.meters
Special Conditions for Safe Use (X):
1. See certificate for special conditions.
Turbine meter terminal
output parameters
Turbine meter terminal
input parameters
V
oc
/U
o
= 2.5 V
V
max
/U
i
= 10 V
I
sc
/I
o
= 253
μ
A
I
max
/I
i
= 1 mA
P
max
/P
o
= 640
μ
W
P
max
/P
i
= 1 mW
C
a
/C
o
= 2.9
μ
F
C
i
= 2.2 nF
L
a
/L
o
= 500 mH
L
i
= 4.7 mH
Summary of Contents for Rosemount 705
Page 2: ......
Page 6: ...vi Reference Manual 00809 0200 4705 Rev AB Contents May 2017 Contents ...
Page 10: ...2 Reference Manual 00809 0200 4705 Rev AB Introduction May 2017 Introduction ...
Page 32: ...28 Reference Manual 00809 0200 4705 Rev AB Commissioning May 2017 Commissioning ...
Page 69: ......