14
Reference Manual
00809-0200-4705, Rev AB
May 2017
Figure 3-4. Antenna Positioning
3.2.2
Installing the transmitter in a remote mount configuration
Figure 3-5. Remote Mount Installation
A. Turbine meter
B. Rosemount 705 Transmitter
C. 1-in. supplied cable gland adapter for turbine meter.
Included:
(1) Cable gland
(1) Cable gland adapter for turbine meter
10 ft.. of cable connection wiring
1. Install the turbine meter or pulse output device according to standard installation practices being
sure to use thread sealant on all of the connections.
2. Pull the cable connection wiring through the supplied cable gland adapter for the turbine meter.
Then pull the cable wiring through the transmitter cable gland.
Note
Pay attention to the orientation of the cable gland to ensure proper connection to transmitter.
3. Attach the turbine meter or pulse output device wiring to the terminals as shown in the wiring
diagrams beginning on
Note
Wireless devices should only be powered up after the Smart Wireless Gateway, in order of proximity from
the Gateway beginning with the closest device. This results in a simpler and faster network installation.
Possible antenna rotation shown.
Antenna rotation allows for best
installation practices in any
configuration.
A
B
C