24
Reference Manual
00809-0100-4570, Rev DA
Electrical Installation
November 2017
Electrical Installation
4.4
Wiring
The units can be connected in different modes and configurations for different external systems such as
PLC or DCS and communications on RS-485, Modbus
®
RTU, and also to RS-485 bus converting adapters
to communication gateways such as Rosemount LinkPro for GMS or GPRS data relay and TCP/IP gateway.
For in-depth details and explanations on wiring and communication, refer to
Communications
Note
If electromagnetic interference is expected, usage of a screened and twisted wired cable is
recommended for the signal lines, which should be connected to the ground reference.
Power supply
For detailed information on power supply specifications, refer to Rosemount 5708
4-20 mA 4-wire: The power supply and signal current inputs must be carried over two separated pairs.
Connection procedure
For detailed information on the connection procedure, see
“Different connection methods” on page 29
Note
Use direct connection between the Rosemount 5708 and the plant (PLC/SCADA).
For a remote connection, use the Rosemount LinkPro remote connection module or a local PC/laptop
computer as detailed in the following drawing.
Always observe the following safety instructions:
Connections must be made only in the complete absence of line voltage.
If over-voltage is expected, overvoltage arresters should be installed.
Use only a safety-certified power supply with dual insulation between the primary and output for
powering the unit. The power supply output rating must be limited to 18-32 Vdc, 1A for a single
Rosemount 5708, and not to be connected to a DC distribution network.
Summary of Contents for Rosemount 5708
Page 1: ...Reference Manual 00809 0100 4570 Rev DA November 2017 Rosemount 5708 3D Solids Scanner ...
Page 2: ...ii Section Title November 2017 Reference Manual 00809 0100 4570 Rev DA ...
Page 8: ...vi Reference Manual 00809 0100 4570 Rev DA Contents November 2017 Contents ...
Page 12: ...2 Reference Manual 00809 0100 4570 Rev DA Introduction November 2017 Introduction ...
Page 202: ...ii Reference Manual 00809 0100 4570 Rev DA Index November 2017 Index ...
Page 203: ...iii Reference Manual 00809 0100 4570 Rev DA November 2017 ...