![Emerson Rosemount 5081FG Reference Manual Download Page 45](http://html.mh-extra.com/html/emerson/rosemount-5081fg/rosemount-5081fg_reference-manual_101186045.webp)
Reference Manual
00809-0100-4882, Rev AB
May 2019
Rosemount 5081FG
3-9
Display Code
Refer to Figure 3-8. This screen is accessible from the CODE screen by
entering 555 and pressing ENTER. The DISPLAY CODE screen identifies the
analyzer access code so you can return to the CODE screen and enter the
code as described in Code. To return to the CODE screen, press NEXT.
Fault Val
Refer to Figure 3-9. Use this screen to set the value that the 4-20 mA output
will drive to and display during a fault condition. Press ENTER to begin
editing. Use the arrow keys to enter a fault value. The fault value can be
between 3.8 and 24 mA. Then, press ENTER to accept the value. Pressing
NEXT displays the UPPER RANGE VAL screen. Refer to Section 5:
Troubleshooting, for the actual fault conditions.
Figure 3-8. Display Code
Figure 3-9. Fault Val
Summary of Contents for Rosemount 5081FG
Page 2: ......
Page 6: ...Reference Manual 00809 0100 4882 Rev AB June 2019 Rosemount 5081FG TOC 4 ...
Page 20: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 1 10 ...
Page 36: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 2 16 ...
Page 53: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 3 17 Figure 3 20 Show Fault ...
Page 82: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 6 6 ...
Page 94: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG A 8 ...
Page 96: ...Reference Manual 00809 0100 4882 Rev AB June 2019 Rosemount 5081FG B 2 ...
Page 98: ...Reference Manual 00809 0100 4882 Rev AB June 2019 Rosemount 5081FG B 2 ...
Page 100: ......
Page 103: ...Index 2 ...