Reference Manual
00809-0100-4882, Rev AB
May 2019
Rosemount 5081FG
2-8
Installing Rosemount
5081 Transmitter
2. The ambient temperature of the transmitter housing must not exceed
65°C (149°F). Locate the electronics in an area where temperature
extremes, vibration, and electromagnetic and radio frequency
interference are minimal.
3. Locate the Rosemount 5081 transmitter within 150 ft (45,7 m) of the
oxygen probe due to wiring and signal considerations.
1. Ensure all components are available to install the Rosemount 5081
transmitter.
2. Choose a method or location to mount the transmitter.
a. Flat Surface Mounting. The transmitter may be mounted on a flat
surface using the threaded mounting holes located on the bottom of
the transmitter housing. Refer to Figure 2-6 for installation
references.
b. Pipe Mounting. An optional pipe mounting bracket is available for
this type of installation. Refer to Figure 2-7 for installation references.
Figure 2-6. Flat Surface Mounting Dimensions
Summary of Contents for Rosemount 5081FG
Page 2: ......
Page 6: ...Reference Manual 00809 0100 4882 Rev AB June 2019 Rosemount 5081FG TOC 4 ...
Page 20: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 1 10 ...
Page 36: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 2 16 ...
Page 53: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 3 17 Figure 3 20 Show Fault ...
Page 82: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 6 6 ...
Page 94: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG A 8 ...
Page 96: ...Reference Manual 00809 0100 4882 Rev AB June 2019 Rosemount 5081FG B 2 ...
Page 98: ...Reference Manual 00809 0100 4882 Rev AB June 2019 Rosemount 5081FG B 2 ...
Page 100: ......
Page 103: ...Index 2 ...