Sensor specifications
(continued)
Table 1-1:
Physical characteristics
Specifications
Temperature (operating)
0 to 50 °C (32 to 122 °F)
Process connection
1 in. MNPT
Wetted parts
Noryl
®
(1)
, Viton, EPDM, Teflon
®
(TFE)
(2)
, and silicone
Cathode
Gold (not normally wetted)
Accuracy
±0.2 ppm at 77 °F (25 °C)
Repeatability
±0.05 ppm at 77 °F (25 °C)
Response time
< 30 sec to 90% of final reading (0 to 2 ppm) at 25 °C (77 °F)
Electrolyte volume
25 mL (approx.)
Electrolyte life
4 to 6 months (approx.)
Cable length (standard integral cable)
25 ft (7.6 m)
Cable length (maximum)
300 ft (91 m)
Sample flow
Flow through
1 to 5 gpm (3.8 to 19 L/min)
Open channel
1 ft/sec (0.3 m/sec)
Low flow cell
2 to 5 gph (7.6 to 19 L/hr)
Agitation produced by bubbles in aeration basins usually
provides adequate flow.
Process connections
1 inch MNPT
Weight/shipping weight
1 lb/3 lb (0.5kg /1.5 kg)
(1) Noryl is a registered trademark of General Electric.
(2) Viton and Teflon are registered trademarks of E.I. duPoint de Nemours & Co.
Plan
2
Rosemount 499ADO
Summary of Contents for Rosemount 499ADO
Page 1: ...Quick Start Guide LIQ QSG 499ADO Rev K July 2017 Rosemount 499ADO Dissolved Oxygen Sensor ...
Page 4: ......
Page 6: ...Contents ii Rosemount 499ADO ...
Page 10: ...Flow through 1 1 2 in Tee Figure 2 2 Install 4 Rosemount 499ADO ...
Page 11: ...Flow through 2 in Tee Figure 2 3 Install Quick Start Guide 5 ...
Page 12: ...Low Flow Cell Figure 2 4 Install 6 Rosemount 499ADO ...
Page 14: ...Install 8 Rosemount 499ADO ...
Page 24: ...Calibrate 18 Rosemount 499ADO ...