Not all calibration procedures should be performed for each Rosemount 4600. Some
procedures are appropriate for bench calibration, but should not be performed during field
calibration.
identifies the recommended calibration procedures for each type of
Rosemount 4600 transmitter for bench or field calibration.
Figure 4-1. Transmitter Data Flow with Calibration Options
Note
Value on PV line should equal the input pressure. Value on AO line should equal the output
device reading.
A. Sensor
B. Input pressure
C. Input device
D. Field Communicator
E. Microprocessor digital PV
F. Transmitter electronics module
G. Analog output
H.20 mA
I. Output device
F
E
A
C
I
H
4600: PT-4763
1
á
Device Setup
Online
2
PV
1000 PSI
3
AO
20.00 mA
4
LRV
0.00 PSI
5
URV
1000 PSI
B
G
D
TRANSMITTER RANGED 0 TO 1000 PSI