2
Install
2.1
Install sensor
If the sensor is installed in a sidestream with the sample draining to open
atmosphere, bubbles may accumulate on the electrodes.
Trapped bubbles will cause errors. Normally, as bubbles accumulate the
conductivity reading drifts down. To control bubble formation, apply a small
amount of back pressure to the sensor.
Figure 2-1: Rosemount 404 contacting conductivity sensor installation
2.2
Electrical installation
For additional wiring information on this product, please refer to the
Table 2-1: Wire color and connections in sensor
Color
Function
Gray
Connects to outer electrode
Clear
Coaxial shield for gray wire
Orange
Connects to inner electrode
Clear
Coaxial shield for orange wire
Red
White with red stripe
White
Clear
Shield for all RTD lead wires
May 2019
Quick Start Guide
Quick Start Guide
7