Note
Do not torque down the transducer retainer to avoid damages to the anti-rotation
pins.
10. Connect the transducer chordset to the transducer retainer. The internal connector
of the transducer chordset is keyed and will only go on one way.
11. Secure the transducer cabling nut by turning clock wise. Ensure the cable nut
threads are correctly aligned.
12. Repeat
for other transducer assemblies if their transducer
capsule assemblies were replaced.
13. Continue with
to use the MeterLink Transducer Swap-out Wizard to
update delta delay time.
3.6
T-200 transducer housing removal and
installation
This section provides the procedures for removing a transducer housing from a T-200
transducer assembly and installing a transducer housing on a transducer stalk assembly.
If the transducer assembly is installed in a meter body, follow the procedure in
to remove the transducer assembly from the meter body
first. After installing or replacing a transducer housing, follow
to install the transducer assembly to the meter body.
WARNING
CONTENTS MAY BE HAZARDOUS
The meter must be fully depressurized and drained before attempting to remove the
T-200 transducer assembly. If gas or fluid begins to leak from the T-200 transducer stalk
assembly, stop immediately and reinstall T-200 stalk assembly.
Failure to comply could cause serious injury or equipment damage.
A. Transducer holder
Maintenance and Troubleshooting Manual
Meter repairs
00809-0600-3104
April 2022
Rosemount Gas Ultrasonic Flow Meters
67