Figure 6-4: Surface Threshold Is Too Low
0
A. Amplitude
B. Distance
C. Disturbing echo misinterpreted as product surface
D. Actual product surface
By adjusting the Surface Threshold the product surface is properly detected as illustrated
in
Figure 6-5: Echo Curve Plot after Surface Threshold Was Adjusted
A. Amplitude
B. Distance
C. After Surface Threshold is adjusted the product surface is correctly detected.
Example 2: Interface peak not found
In interface applications where the bottom product has a relatively low dielectric constant
(<40), or if the signal is attenuated in the upper product, the amplitude of the reflected
signal from the interface is relatively low and difficult for the transmitter to detect. In such
a case it may be possible to detect the reflected signal from the interface if the Interface
Threshold is adjusted.