information about the required IO-Link hardware and software are available at
3.3
Adjustable parameters
Use IO-Link software to configure these parameters.
Parameter
Description
OU2
Output function: analog signal is 4-20 mA (I) or 20-4 mA (Ineg)
ASP2
Analog start point for the measured temperature value.
AEP2
Analog end point for the measured temperature value. The minimum
difference between ASP2 and AEP2 is 9 °F (5 °C).
COF
Zero point calibration. Offsets the zero point ± 18 °F in steps of .18 °F (±
10 °C in steps of 0.1 °C).
FOU2
Behavior of the output after a fault.
•
ON: the analog signal is driven to 21.1 mA.
•
OFF: the analog signal is driven to 3.5 mA.
Uni
Sets the temperature unit in °F or °C.
Update rate of display
The rate at which the transmitter updates the displayed temperature
measurement.
•
d1: 50 ms
•
d2: 200 ms
•
d3: 600 ms
Orientation of display
rd: rotates the display by 180°.
3.4
Operating modes
The temperature transmitter uses different operating modes depending on the measured
temperature.
After turning on the temperature transmitter the temperature transmitter displays the
current measured temperature and the analog output provides a signal proportional to
the measured temperature.
If the temperature is outside of the measuring range the temperature transmitter displays
OL or UL.
Operation
Quick Start Guide
January 2019
00825-0100-4328
10
Quick Start Guide
Summary of Contents for Rosemount 326T
Page 1: ...Quick Start Guide 00825 0100 4328 Rev 01 January 2019 Rosemount 326T Temperature Transmitter ...
Page 2: ...2 ...
Page 4: ...Contents Quick Start Guide January 2019 00825 0100 4328 4 Quick Start Guide ...
Page 12: ...Temperature resistance Quick Start Guide January 2019 00825 0100 4328 12 Quick Start Guide ...
Page 14: ...Product certifications Quick Start Guide January 2019 00825 0100 4328 14 Quick Start Guide ...